Technical data
68
10.2 Output
10.2.1 Output signal
Relay, transmitter power supply (standard)
Current, voltage, pulse, intrinsically safe transmitter power supply (option)
10.2.2 Signal on alarm
No measured value visible on the LC display, no background illumination, no sensor power
supply, no output signals, relays behave in safety-oriented manner.
10.2.3 Current/voltage output
Span:
0/4 to 20 mA (active), 0 to 10 V (active)
Load:
600 Ω (current output)
Max. loop current 22 mA (voltage output)
Signal characterization:
Signal freely scalable
Galvanic isolation towards all other circuits
10.2.4 Pulse output
Frequency range to 12.5 kHz
•I
max
= 200 mA
•U
max
= 28 V
•U
low/max
= 2 V at 200 mA
Pulse width = 0.04 to 2000 ms
10.2.5 Relay
Signal characterization:
Binary, switches when the limit value is reached
Switch function: limit relay switches for the operating modes:
Minimum/maximum safety
Alternating pump control function
Batch function
Time control
Window function
Gradient
Device malfunction
Sensor malfunction
Switching threshold:
Freely programmable
Vista de pagina 67